- 1 Robotic hand.
- 2 Pipetting probe, 1ul stepping (1000ul volumes).
- Total 25 racks, arbitrary rack can be defined as sample or reagent on the software.
- 4 units 96 well microplates (pipetting,incubator,reading).
- 1 unit independent washer &reader( two 8 nozzle manifolds).
FAE-4 Fully automated ELISA processor
QUICK OVERVIEW
Product Code: 845
Brand: Elisa System
FAE-4
Share this product on social networks
Specification
Model | FAE-4 | FAE-4 | |
---|---|---|---|
Sample pipetting unit | Sample rack | 24 sample tubes/rack, at most 25 racks | |
Primary sample tubes | 11.5-13mm | ||
Pipetting | 8ul~1000ul, (10ul) CV<5% | ||
Time to dispense | 15 minutes (typical)^100ul of 96 samples to plate from sample tubes or dilution tubes. | ||
Dispensing system | 1 aspirating-dispensing probe with X-Y-Z movement ;dilution systemwith a syringe of 1000ul, continuous flow washing station. | ||
Reagent pipetting unit | Reagent racks | 2 reagent tanks(60ml)/ rack, at most 25 racks | |
QC racks | 24 positions/QC rack, at most 25 racks | ||
Pipetting | 8ul~1000ul,1ul stepping, | ||
pipetting precision | (100ul )CV≤0.5% | ||
Time to dispense | 2 minutes(typical)^100ul of 96 reagents to plate from reagent bottle | ||
Washing unit | Programmable | Yes(options & parameters) | |
Wash containers | 4 wash buffers at 2.0L, with level-sensing | ||
Waste containers | 10L with waste full sensor | ||
Washing system | two 8 nozzle manifolds, one to dispense and the other one to aspirate; | ||
Washing residual | <2μl | ||
Reading unit | Reading system | 8 independent photometric channels with mono and bichromatic reading capability | |
Photometric range | 0.000 ~3.000 OD | ||
Spectral range | 400-700nm | ||
Optical filters | 2 precision filters (450and 630nm); 2 more on request (4 filters max, 405 nm, 450 nm, 492 nm and 630 nm) | ||
Reading inaccuracy | inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD | ||
Incubation unit | Incubators | 4 independent units | |
Programmable | Yes | ||
Temperature range | RT to 45 °C | ||
Temperature accuracy | ±0.5℃ | ||
Temperature staility | ±0.5℃ | ||
Software unit | Software system | Windows XP or Windows 7 or above, LIS system available, realize two-way data communications | |
Function application | work list set up, patient reports and result filing; calibration curve, storage for each carried out test | ||
Data reduction | cut-off (qualitative) | ||
Physical Specification | Power Supply | AC220V+ 10%,50/60Hz,450W | |
Instrument Dimension | 110*72*70cm (L*W*H) | ||
Instrument Weight | 100kg | ||
Package Dimension | 138*91*98cm(L*W*H) | ||
Total Weight | 150kg |