Blood Glucose Test Strip (hereinafter referred to as Blood Glucose Strip or Strip) adopts the new biosensor technology and siphon principle.
Glucose oxidase and conductive medium is immobilized in the surface of the electrode, when the glucose in the blood contacts with the enzyme on the electrode
and the conductive medium, specific reaction will be initiated and instantaneous current will be produced, the size of current is directly proportional to the glucose
concentration, and through the Blood Glucose Meter, the current signal is converted into blood glucose value, which is displayed on the screen.
Technical parameter
Technical parameter
|
|
Model |
BGM30 |
Blood sample |
Capillary whole blood |
Dimension |
8.5cm×5.4cm×1.9cm |
Volume of blood |
3ul |
Memory results |
500 |
Measuring range |
1.1-33.3mmol/L |
Measuring time |
10 seconds |
Testing temperature |
5℃~42℃(41℉~107.6℉) |
Testing humidity |
≤80% |
Display |
LCD |
Weight |
50g with battery |
Battery life |
1000 times |
Power supply |
3-volt button cell |
Features
|
|
Test sample |
fresh capillary whole blood |
Type of sample used for calibration |
Venous plasma |
Blood volume |
about 3µl |
Measurement unit |
mmol/L and mg/dL, switchable |
Instrument calibration |
automatic calibration with test piece code |
Calculation of average value |
7, 14, 21 and 28 days |
Display data and time |
12-hour system |
Storage and transport temperature |
-20℃~55℃ (-4℉~131℉) |
Storage and transport relative humidity |
not higher than 85% |
Equipment rating |
3Vdc, 3mA (one CR2032 battery) |
Quality control solution |
use key to isolate quality control value to exclude it |
Alarm setting |
add alarm function |
Bloodsucking symbol display |
display on the LCD screen |